Friday, August 31, 2012

Gas Chromatography/Mass Spectroscopy (1)


DESCRIPTION OF TECHNIQUE

Gas chromatography / mass spectrometry (GC/MS) is the marriage of two analytical methods into a versatile technique for the identification of complex volatile materials. Gas chromatography (GC) effectively separates the different constituents of the sample for subsequent analysis and identification by mass spectrometry (MS).

The chromatographic separation relies on the interaction of the sample with a mobile phase and a stationary phase within the GC instrument column. The sample is carried through the column by the mobile phase, typically an inert gas. However, the sample is slowed in its travel through the column as the sample molecules repeatedly adsorb and desorb from the stationary phase in the column. The affinity of a particular molecule for the stationary phase determines the retention time of that constituent in the column. The molecules for each component of the sample will travel through the column at nearly the same rate and exit (elute) from the column within a narrow time band that is specific to that component. Thus, compounds with different retention times in the column are physically separated for presentation to a detector and analyzer.

The typical GC capillary column consists of a smalldiameter tube with a thin film of a high-molecularweight polymer coated on the inside. The polymer is the stationary phase for the chromatographic process. The mobile phase can be any inert gas, but is typically helium. The instrument also includes a heated injection port to vaporize all volatile constituents of the sample and an oven to keep the constituents in gas form as they pass through the column.

As a sample constituent elutes from the GC column, it enters the ionization chamber of the mass spectrometer where the molecules are ionized, typically by electron impact. When an electron impact with a sample molecule results in the loss of an electron from the molecule, a positive ion is formed. Some of the molecular ions are further fragmented into daughter ions and neutral fragments. The positive ions are then repelled out of the ionization chamber by a small positive charge within the chamber. Negative ions are also formed by the electron impact, but are not analyzed. The positive ions are separated according to their mass by a mass analyzer. The mass analyzer most commonly used in GC/MS is the quadrapole filter, in which the ions pass by four hyperbolic magnetic poles created by a radio frequency field. The magnetic poles separate the ions by their mass/charge ratio, successively focusing ions with increasing mass onto a detector for counting. The analyzer scans step-wise through a set range of mass values to evaluate the relative abundance of ions at each mass value. The quadrapole filter can perform a complete mass scan within the duration of a single GC elution band.

No comments:

Post a Comment